MythoVentT
Our latest product, the MythoVentT, complements our current portfolio of ventilators, MythoVent and MythoVent2. The MythoVentT is a turbine driven, modern device that meets the needs of neonatal, pediatric and adult patients, and provides maximum reliability, low operating cost and ease of use.
MythoVentT Key Features
- Turbine operated
- 12" Touch screen
- Patient connected flow sensor for neonates
- Mainstream CO2 (optional)
Ventilation modes |
Pressure | CPAP, BILEVEL, SIMV, PRVC, a/c |
Volume | MMV, SIMV, a/c |
ASB (Assisted Spontaneous Breathing) | CPAP, BILEVEL, SIMV, MMV |
ABV (Apnea Backup Ventilation) | CPAP |
Additional | NIV, TC, Sigh |
Manual | Single breath, inspiratory / expiratory hold, ventilation pause automatic |
Advanced |
CO2 capnography | Optional |
Ventilation settings |
Ventilation frequency / ABV breath frequency | approx. 1 to 120 per minute |
Tidal volume / Volume limit / ABV volume limit | approx. 20 to 1500 mL |
Minute volume | approx. 0.5 to 30 L/min |
Inspiratory flow | approx. 0 to 60 L/min |
Inspiratory pressure / ABV pressure / Pressure limit | approx. 5 to 50 mbar |
PEEP | 0 to 30 mbar |
Inspiratory time | approx. 0.1 to 15 s |
I:E ratio / ABV I:E ratio | approx. 1:6 to 6:1 |
Inspiratory O2concentration | approx. 21 to 100 % |
Pressure support | approx. 0 to 30 mbar |
PHIGH (BILEVEL only) | approx. 5 to 30 mbar |
THIGH (BILEVEL only) | approx. 1 to 15 s |
PLOW (BILEVEL only) | approx. 0 to 25 mbar |
TLOW (BILEVEL only) | approx. 1 to 45 s |
Upper pressure limit (PRVC only) | approx. 5 to 50 mbar |
Lower pressure limit (PRVC only) | approx. 2 to 30 mbar |
Advanced ventilation settings |
Safety pressure | approx. 30 to 50 mbar |
Apnea time | approx. 5 to 60 s |
Max. inspiratory time for pressure support | approx. 0.5 to 5 s |
Inspiratory pressure trigger | approx. -10 to -1 mbar |
Inspiratory flow trigger | approx. -0.5 to 20 L/min |
Expiratory trigger | approx. 5 to 90 % |
Sigh offset pressure | approx. 5 to 20 mbar |
Sigh rate | approx. 1/10 to 1/200 |
NIV leakage compensation | approx. 5 to 50 %, Auto |
Values |
Ppeak | Peak pressure |
PEEP | PEEP |
Pmean | Mean pressure |
Pplat | Plateau pressure |
Fi | Inspiratory flow |
Fe | Expiratory flow |
Vi | Inspiratory volume |
Ve | Expiratory volume |
MVi | Inspiratory minute volume |
MVe | Expiratory minute volume |
f | Breathing frequency |
I:E | I:E ratio |
Ti | Inspiratory time |
Te | Expiratory time |
R | Resistance |
C | Compliance |
PEEPi | Intrinsic PEEP |
Rsbi | Rapid swallow breathing index |
O2i | Inspiratory O2 concentration |
Curves |
Pressure | Inspiratory pressure and PEEP (realtime) |
Flow | Inspiratory and expiratory flow (realtime) |
Volume | Inspiratory and expiratory volume (realtime) |
Loops |
P/V | Pressure / Volume loop (realtime) |
Flow/V | Flow / Volume loop (realtime) |
Trend |
Parameters | Ppeak, PEEP, Pmean, Pplat, Flow, Vtid, MV, f, R, C, O2 |
Time scale | 6 to 96 h |
Display | up to 6 trends and alarm-line simultaneously |
Adjustable alarms |
Ppeak | High / Low |
PEEP | High / Low |
Pmean | High / Low |
Flow | High / Low |
Vtid | High / Low |
MV | High / Low |
f | High / Low |
O2 | High / Low, always active |
Advanced alarm settings |
Limits auto | Set limits according a 30 minute trend |
Limits standard | Reset limits to standard values |
Alarms off | Silence alarms for 2 minutes |
Alarm list | Show list of the 10 last alarms |
Alarm sound | Set Alarm frequency and loudness |
Additional alarms |
Apnea | No spontaneous breaths within set time |
Battery powered | Loss off mains |
Battery | At 5, 10 and 15 % |
Device check necessary | Information of incomplete device check |
Technical alarms | Technical alarms and warnings |
Event-Log | Log for 5000 entries with date and time |
Display and Connectors |
Screen | 12.1" TFT color display, up to 6 curves, 3 loops and 24 values simultaneously |
Brightness | Auto, dark, medium, bright |
Connectors | RS232 for service, grounding, Ethernet (optional), USB (optional), nurse call (optional) |
Energy and Gas supply |
Mains power connection | 100 to 240 VAC, 50 / 60 Hz |
Power consumption | approx. 150 W |
Internal battery | 12 VDC, approx. 30 minutes of ventilation |
Air gas | turbine-powered, non-condensing gas |
O2 gas | 2.8 bar to 6 bar, dry gas |