Kamis, 16 Juni 2011

MythoVentT Ventilator


MythoVentT

Our latest product, the MythoVentT, complements our current portfolio of ventilators, MythoVent and MythoVent2. The MythoVentT is a turbine driven, modern device that meets the needs of neonatal, pediatric and adult patients, and provides maximum reliability, low operating cost and ease of use.

MythoVentT Key Features

  • Turbine operated
  • 12" Touch screen
  • Patient connected flow sensor for neonates
  • Mainstream CO2 (optional)






Ventilation modes

PressureCPAP, BILEVEL, SIMV, PRVC, a/c
VolumeMMV, SIMV, a/c
ASB (Assisted Spontaneous Breathing)CPAP, BILEVEL, SIMV, MMV
ABV (Apnea Backup Ventilation)CPAP
AdditionalNIV, TC, Sigh
ManualSingle breath, inspiratory / expiratory hold, ventilation pause automatic

Advanced

CO2 capnographyOptional

Ventilation settings

Ventilation frequency / ABV breath frequencyapprox. 1 to 120 per minute
Tidal volume / Volume limit / ABV volume limitapprox. 20 to 1500 mL
Minute volumeapprox. 0.5 to 30 L/min
Inspiratory flowapprox. 0 to 60 L/min
Inspiratory pressure / ABV pressure / Pressure limitapprox. 5 to 50 mbar
PEEP0 to 30 mbar
Inspiratory timeapprox. 0.1 to 15 s
I:E ratio / ABV I:E ratioapprox. 1:6 to 6:1
Inspiratory O2concentrationapprox. 21 to 100 %
Pressure supportapprox. 0 to 30 mbar
PHIGH (BILEVEL only)approx. 5 to 30 mbar
THIGH (BILEVEL only)approx. 1 to 15 s
PLOW (BILEVEL only)approx. 0 to 25 mbar
TLOW (BILEVEL only)approx. 1 to 45 s
Upper pressure limit (PRVC only)approx. 5 to 50 mbar
Lower pressure limit (PRVC only)approx. 2 to 30 mbar

Advanced ventilation settings

Safety pressureapprox. 30 to 50 mbar
Apnea timeapprox. 5 to 60 s
Max. inspiratory time for pressure supportapprox. 0.5 to 5 s
Inspiratory pressure triggerapprox. -10 to -1 mbar
Inspiratory flow triggerapprox. -0.5 to 20 L/min
Expiratory triggerapprox. 5 to 90 %
Sigh offset pressureapprox. 5 to 20 mbar
Sigh rateapprox. 1/10 to 1/200
NIV leakage compensationapprox. 5 to 50 %, Auto

Values

PpeakPeak pressure
PEEPPEEP
PmeanMean pressure
PplatPlateau pressure
FiInspiratory flow
FeExpiratory flow
ViInspiratory volume
VeExpiratory volume
MViInspiratory minute volume
MVeExpiratory minute volume
fBreathing frequency
I:EI:E ratio
TiInspiratory time
TeExpiratory time
RResistance
CCompliance
PEEPiIntrinsic PEEP
RsbiRapid swallow breathing index
O2iInspiratory O2 concentration

Curves

PressureInspiratory pressure and PEEP (realtime)
FlowInspiratory and expiratory flow (realtime)
VolumeInspiratory and expiratory volume (realtime)

Loops

P/VPressure / Volume loop (realtime)
Flow/VFlow / Volume loop (realtime)

Trend

ParametersPpeak, PEEP, Pmean, Pplat, Flow, Vtid, MV, f, R, C, O2
Time scale6 to 96 h
Displayup to 6 trends and alarm-line simultaneously

Adjustable alarms

PpeakHigh / Low
PEEPHigh / Low
PmeanHigh / Low
FlowHigh / Low
VtidHigh / Low
MVHigh / Low
fHigh / Low
O2High / Low, always active

Advanced alarm settings

Limits autoSet limits according a 30 minute trend
Limits standardReset limits to standard values
Alarms offSilence alarms for 2 minutes
Alarm listShow list of the 10 last alarms
Alarm soundSet Alarm frequency and loudness

Additional alarms

ApneaNo spontaneous breaths within set time
Battery poweredLoss off mains
BatteryAt 5, 10 and 15 %
Device check necessaryInformation of incomplete device check
Technical alarmsTechnical alarms and warnings
Event-LogLog for 5000 entries with date and time

Display and Connectors

Screen12.1" TFT color display, up to 6 curves, 3 loops and 24 values simultaneously
BrightnessAuto, dark, medium, bright
ConnectorsRS232 for service, grounding, Ethernet (optional), USB (optional), nurse call (optional)

Energy and Gas supply

Mains power connection100 to 240 VAC, 50 / 60 Hz
Power consumptionapprox. 150 W
Internal battery12 VDC, approx. 30 minutes of ventilation
Air gasturbine-powered, non-condensing gas
O2 gas2.8 bar to 6 bar, dry gas